A regularized quantitative ultrasound method for simultaneous calculation of backscatter and attenuation coefficients
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
SPIE
Acceso al texto completo solo para la Comunidad PUCP
Abstract
Quantitative ultrasound has shown potential of improving medical diagnosis. In this work, a regularized power law (RPL) for the joint estimation of backscatter coefficient (BSC) and attenuation coefficient (AC) parameters was derived and tested with simulated phantoms and in vivo. For the RPL method, a total variation regularization term regarding the BSC and AC parameters were used. The results were compared with ground truth simulated values. An improvement of precision can be appreciated without compromising the accuracy by factor of 80% and 54% according to bias and coefficient of variation, respectively. The in vivo experiments showed comparable results of the algorithm with the literature (i.e. fibroadenoma:1.90 ± 0.31dB.cm−1 .MHz−1 , normal tissue: 0.62 ± 0.20dB.cm−1 .MHz−1 ). The results suggest the RPL method has the potential to accurately and precisely estimating BSCs and ACs.
Description
Keywords
Attenuation, Attenuation coefficient, Coefficient of variation, Backscatter (email), Ultrasound, Physics, Computational physics, Regularization (linguistics), Nuclear magnetic resonance, Materials science, Optics, Mathematics, Statistics, Acoustics, Computer science, Telecommunications