In this research and development, we present the design and manufacture of a radon chamber
(PUCP radon chamber), a necessary tool for the calibration of passive detectors, verification
of the operation of active radon monitors as well as diffusion chamber calibration used in
radon measurements in air, and soils. The first chapter is an introduction to describe radon
gas and national levels of radon concentration given by many organizations. Parameters that
influence the calibration factor of the LR 115 type 2 film detector are studied, such as the
energy window, critical angle, and effective volumes. Those are strongly related to the etching
processes and counting of tracks all seen from a semi-empirical approach studied in the second
chapter. The third chapter presents a review of some radon chambers that have been reported
in the literature, based on their size and mode of operation as well as the radon source they use.
The design and construction of the radon chamber are presented, use of uranium ore (autunite)
as a chamber source is also discussed. In chapter fourth, radon chamber characterization
is presented through leakage lambda, homogeneity of radon concentration, regimes-operation
modes, and the saturation concentrations that can be reached. Procedures and methodology
used in this work are contained in the fifth chapter and also some uses and applications of the
PUCP radon chamber are presented; the calibration of cylindrical metallic diffusion chamber
based on CR-39 chips detectors taking into account overlapping effect; transmission factors of
gaps and pinhole for the same diffusion chambers are determined; permeability of glass fiber
filter for 222Rn is obtained after reach equilibrium through Ramachandran model and taking
into account a partition function as the rate of track density. The results of this research have
been published in indexed journals. Finally, the conclusion and recommendations that reflect
the fulfillment aims of this thesis are presented.